Description
Quick Details
1.Scanning mode: Convex/Linear/Micro-convex/Trans-rectal;
2.Display modes: B, B+B, B+M, M, 4B
3.Scanning depth: 240mm
4.Gray scales: 256
5.Function of image magnification: image zoom (real-time status under B mode), up to 5 times (area); local amplification function which can be enlarged under the real-time and froze
6.Image conversion: up/down, left/right, black/white
7.Function of puncture guide
Packaging & Delivery
Packaging detail:Standard export package Delivery detail:within 7-10 workdays after receipt of payment |
Specifications
Specification of Laptop Animal Ultrasound Scanner MSLPU56V:
1.Scanning mode: Convex/Linear/Micro-convex/Trans-rectal;
2.Display modes: B, B+B, B+M, M, 4B
3.Scanning depth: 240mm
4.Gray scales: 256
5.Function of image magnification: image zoom (real-time status under B mode), up to 5 times (area); local amplification function which can be enlarged under the real-time and froze
6.Image conversion: up/down, left/right, black/white
7.Function of puncture guide
Standard Configuration of Laptop Animal Ultrasound Scanner MSLPU56V:
12 inches LCD main unit 1set;
60R/3.5MHZ Convex array probe(2.0/3.5/4.5/5.0/5.5/6.5/7.5MHz) 1pc ;
Power line 1pc;
Ground line 1pc;
Fuse 4 pc;
User’s manual in English 1pc;
Warranty card 1pc;
Options & Accessories of Laptop Animal Ultrasound Scanner MSLPU56V
7.5MHZ Linear array probe (7.5/8.5/10.5MHz);
5.0MHZ Micro-Convex array probe (4.0/5.0/5.5MHz);
7.5MHZ Trans-rectal array probe (5.0/7.5/10.0MHz);
Mitisubishi Video Printer P93C;
Clent Usage Photos of Laptop Animal Ultrasound Scanner MSLPU56V
If you want to know more details, please contact us.
Medial & Video Of of Laptop Animal Ultrasound Scanner MSLPU56V
If you want to know more details, please contact us.
Technical Parameter of Laptop Animal Ultrasound Scanner MSLPU56V
Max dynamic range>=100db;
Detecting depth>=190mm, scanning depth>=240mm
The geometric location accuracy: vertical<=5%; lateral<=5%; blind spot: <=4mm;
Vertical resolution: <=1mm (depth<=130mm) < =2mm (130mm<depth<170mm);
Lateral resolution: <=1mm (depth<=130mm) < =2mm (130mm<depth<160mm);
Gamma: 4 gray scale γ transform function;
Variety of focus selection and multi-stage dynamic focusing with 4 transmit focus; support to launch multi-focus combination;
Built-in pseudo color processor, Pal-D and VGA video output, can connect external screen monitors, color monitors, VCRs, video image printer and other equipment.
Fixed characters and real-time clock and calendar display;
Software reset function;
Printer function: can connect laser printer and Video printer directly;
Comment: Date, Name, Sex, Age, Hospital, Annotation;
Zooming: *1.0, *1.2, *1.5, *2.0;
Port: USB 2.0, VGA, Video connector;
USB Function of image store: more than Sixteen thousand images;
Power supply: AC 110V-240V, 50HZ/60HZ;
Probe connector: 2;
USB/VGA/ ports;
Video printer connector;