Hot Sale Quantitative Electrolyte Analyzer MSLEA06
Intended Use: The Na+, K+, Cl– ,Ca2+ methods on the CBS Electrolyte Analyzer system are in Vitro diagnostic tests intended for the quantitative measurement of Sodium, Potassium, Ionized Calcium and Chloride in serum and Urine. They are also intended for the quantitative measurement of Na+ ,K+, Cl– , Ca2+ in diluted urine.
Summary: The Sodium, Potassium, Ionized Calcium and Chloride methods use direct sensing Multi-Biosensor Module (MBM) to develop an electrical potential proportional to the activity of each specific ion in the sample.
Hot Sale Quantitative Electrolyte Analyzer MSLEA06
Hot Sale Quantitative Electrolyte Analyzer MSLEA06
Principles of Procedure: CBS multi-sensor has four ion selective electrodes ( ISE) measuring ion concentration of sodium, potassium, Ionized Calcium and chloride in sample. Another chloride electrode is integrated in cartridge as a reference electrode. Each ion selective membrane undergoes a specific reaction with the ion contained in the sample. The membrane reacts to the difference of potential of sample and the standard solution A and calculation. Then ion concentration in the sample is determined.
Installation: Open the bag, put module into the box, close the cover . Installation details please refer to "MSLEA06 Electrolyte Analyzer Manual."
Sample Collection: The samples collection of serum and urine by routine method can be used in CBS-Multi-sensor.
1. Plasma samples by rapid centrifugation of heparinized whole blood, were isolated from the red blood cells, and seal the sample in a test tube;
2. Serum samples collected in a blood collection cup, 30 minutes and centrifuged to remove serum clot, sealed tube;
3. Urine samples were collected at random in accordance with statutes or 24-hour urine.
Specification
Precautions: Module only for in vitro diagnostic. from the cold storage environment should keep 1 hour at room temperature before installation. Used Modules are Medical waste maybe contain human body fluids with infectious agents. care to avoid skin contact and ingestion. should be properly handled in accordance with relevant rules and regulations.
Use Limitations: the serum, plasma, urine, cerebrospinal fluid, pleural effusion, ascites, electrolyte ions (potassium, sodium, chloride, calcium).
Whole blood specimens cannot be analyzed. The interaction of acidifying reagent and whole blood sample forms precipitates which may block the flow lines.
Storage condition: Advises to store the multi-biosensor at room temperature, the period of validity is 12 months.
Expiration: Valid for 1 year, after opening the package is valid for 1 month
conditions of use: At room temperature of 15~30 C, the service life is no less than 600 samples, or 10 samples per day, the number of days used is no less than 20 days.