Anesthesia machine > Anesthesia Ventilator

Advanced Anesthesia Ventilator for sale-EKONGA15

MODEL / EKONGA15

This anesthesia machine consists of a main unit, an anesthetic vaporizer, a flow meter, an anesthesia respirator and a respiratory circuit system.

  • This EKONGA15 Anesthesia Machine is an essential important anesthesia device in an operation room. Its function is
  • This EKONGA15 Anesthesia Machine is equipped with an accurate dedicated anaesthetic vaporizer and a safety device
  • for preventing cyanosis and necessary alarm system. During the anesthesia, the patient’s respiratory functions can be
  • 4 For oxygen and laughing gas, a special pressure regulator with a safety valve will be provided. The exhaust pressure
PRODUCT OVERVIEW

Product Details

Quick Details

This anesthesia machine consists of a main unit, an anesthetic vaporizer, a flow meter, an anesthesia respirator and a respiratory circuit system.

Specifications

Advanced Anesthesia Ventilator for sale-EKONGA15

Advanced Anesthesia Ventilator EKONGA15

MSL Anesthesia Ventilator machine - EKONGA15 Overview

This EKONGA15 Anesthesia Machine is an essential important anesthesia device in an operation room. Its function is

to provide oxygen and anesthetic agent to a patient needing to go through an anesthesia operation and perform

ventilation management.

This anesthesia machine is intended for use in children and adults with a body weight heavier than 10kg.

This EKONGA15 Anesthesia Machine is equipped with an accurate dedicated anaesthetic vaporizer and a safety device

for preventing cyanosis and necessary alarm system. During the anesthesia, the patient’s respiratory functions can be

managed by using a microcomputer controlled pneumatic electrically controlled synchronizing anesthesia respirator.

Each connection part of the whole machine is a standard interface. A highly efficient and large volume soda lime absorber

can minimize the re-inhalation of carbon dioxide by the patient.

advanced Anesthesia Ventilator
Advanced Anesthesia Ventilator EKONGA15 for sale
new Anesthesia Ventilator EKONGA15
cheap Anesthesia Ventilator EKONGA15
best Anesthesia Ventilator EKONGA15

Operating conditions for EKONGA15 Anesthesia Machin e:

---------Ambient temperature: 10 ~ 40 ℃;

---------Relative humidity: no higher than 80 %;

--------Atmospheric pressure: 860 hPa ~ 1060 hPa.

--------Power requirement: ~220-240V, 50/60Hz, 40VA, to be well grounded.

--------Air source requirement: medical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa.

Attention: the AC power supply used for the anesthesia machine must be well grounded.

Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:

1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with

51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of

Anesthesia System.

MSL New Anesthesia Ventilator for sale

Cheap Anesthesia Ventilator EKONGA15 Structural Characteristics and Operating Principles

This anesthesia machine consists of a main unit, an anesthetic vaporizer, a flow meter, an anesthesia respirator and a

respiratory circuit system.

New Anesthesia Ventilator

Best Anesthesia Ventilator for sale-EKONGA15 Technical

1 Operating mode: circulatory closed, semi-closed and semi-open.

2 Gas requirement: medical oxygen and laughing gas with a pressure ranging from 0.3MPa to 0.5 MPa.

3 The maximum error of the pressure testing device should not exceed ± (4% of the full scale reading + 8% of the

actual reading).

4 For oxygen and laughing gas, a special pressure regulator with a safety valve will be provided. The exhaust pressure

of the safety valve should not be higher than 6 kPa.

5 Indication range of the oxygen and laughing gas flow meter: 0.1 L/min ~ 10 L/min.

When the flow rate ranges from 10% of the full scale value to 100%, the scale accuracy should be within ±10%

of the indicated value.

6. The flow meter is equipped with an oxygen-laughing gas proportional control device.

When the oxygen concentration in the mixed gas N2O/O2 conveyed by the anesthesia

machine is no lower than 20% (V/V) or the FiO2 is lower than 20%, the machine will emit an

alarm.

7. When the oxygen pressure of the anesthesia machine is 0.20MPa±0.05MPa, the machine raises a low gas source

pressure alarm which is a high priority alarm, and shuts off the laughing gas conveyed to the common gas outlet.

8. Oxygen Flush:25~75 L/Min;

9. Anesthetic gas concentration adjustment range of vaporizer: 0~5%, relative error ±20 %.

10. The exhaust pressure of the safety valve of the respiratory circuit is no higher than 6 kPa.

11. Anesthesia ventilator

11.1 Respiration Mode:IPPV,SIPPV,Manu

11.2 Respiration Frequency:4~40bpm

11.3 I/E Ratio:1:1.5~1:4

11.4 Tidal Volume:50~1500mL

11.5 Ptr:-10~10hPa

11.6 Controlled ventilation and assisted ventilation switching time: 6s

11.7 Maximum safety pressure: ≤ 12.5 kPa.

11.8 Pressure limit range: 1~6 kPa

11.9 Airway pressure alarm: adjustment range of upper alarm limit: 0.3kPa ~ 6 kPa, permissible error ±0.2 kPa, or ±15%

(whichever is greater), the machine should immediately raise a high level alarm when the airway pressure rises to the alarm

value; the lower alarm limit ranges from 0.2 to 5 kPa and the permissible error is ±0.2 kPa or ±15% (whichever is greater).

The machine should immediately raise a medium level alarm when the airway pressure drops to the alarm value and

such condition lasts 4 to 15s.

3.11.10 Tidal volume alarm: the upper alarm limit ranges from 50 to 2000ml, the permissible error is ±20%,

the adjustment range of the lower alarm limit is 0~1800ml, the permissible error is ±20% and the machine should raise

a medium level alarm.

3.11.11 Ventilation volume lower alarm limit: the adjustment range is 0~12 L/min, and the permissible error is ±20%.

The ventilation volume upper alarm limit is fixed to be 25 L/min, the permissible error is ±20% and the machine should

raise a medium level alarm.

3.11.12 Power supply fault alarm: the machine should give audible alarm in case of power outage and the alarm should

last more than 120s.

3.11.13 The silencing time of the audible alarm signal should be less than 120s. The alarm status should not be

deactivated

and visual alarm signal should not be intermittent.

3.11.14 The emergency battery should be a lead acid battery with a rated output voltage of DC 12V. The duration of the

operation of the anesthetic respirator driven by the battery should be more than 60min.

Anesthesia Ventilator for sale
MSL New Anesthesia Ventilator
MSL New Anesthesia Ventilator for sale
TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

-------Ambient temperature
10 ~ 40 ℃
-------Relative humidity
no higher than 80 %
------Atmospheric pressure
860 hPa ~ 1060 hPa.
------Power requirement
~220-240V, 50/60Hz, 40VA, to be well grounded.
------Air source requirement
medical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa.
Attention
the AC power supply used for the anesthesia machine must be well grounded.
93, an oxygen monitor complying with ISO 7767
1997 and an expiratory gas volume monitor complying with
1.4.2 of Medical Electrical Equipment Part II
Special Requirements for Safety and Basic Performance of
Operating mode
circulatory closed, semi-closed and semi-open.
Gas requirement
medical oxygen and laughing gas with a pressure ranging from 0.3MPa to 0.5 MPa.
Indication range of the oxygen and laughing gas flow meter
0.1 L/min ~ 10 L/min.
Oxygen Flush
25~75 L/Min
Anesthetic gas concentration adjustment range of vaporizer
0~5%, relative error ±20 %.
Respiration Mode
IPPV,SIPPV,Manu
Respiration Frequency
4~40bpm
I/E Ratio
1:1.5~1:4
Tidal Volume
50~1500mL
Ptr
-10~10hPa

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.