Patient monitor

Display screen vet patient monitor for sale EKONMP06

MODEL / EKONMP06

Display screen vet patient monitor for sale EKONMP06. Plug‐in rechargeable battery, AC/DC available.

  • 1" color TFT screen
  • Anti‐defibrillation design
  • Supporting wire or wireless network.
  • Accuracy ±1% or±1bpm, which is greater
PRODUCT OVERVIEW

Product Details

Quick Details

12.1" color TFT screen

Specifications

Display screen vet patient monitor for sale EKONMP06

Professional vet patient monitor for sale EKONMP06

12.1" color TFT screen

Multi‐channel display

Big font interface

ST and arrhythmia analysis

Drug dose calculation and titration table

OxyCRG dynamic view display

Respiration monitoring with apnea alarms

Alarm review, arrhythmia review

Patient info. Input

Anti‐defibrillation design

NIBP over pressure protection

Plug‐in rechargeable battery, AC/DC available.

Supporting wire or wireless network.

use vet patient monitor for sale EKONMP06

MSL vet patient monitor for sale EKONMP06 Technical Specifications

ECG

Lead Mode 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)

Lead selection I, II, III, avR, avL, avF, V

Wavef orm 2 ch

Lead mode 3 Leads ( R, L, F or RA, LA, LL)

Lead selection I, II, III,

Wave form 1 ch

Gain 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto

HR and Alarm Range Adult 15 ~ 300 bpm

Neo/Ped 15 ~ 350 bpm

Accuracy ±1% or±1bpm, which is greater

Resolution 1 bpm

Sensitivity > 200 (uVp-p)

Differential Input Impedance > 5 MΩ

CMRR Monitor > 105 dB

Operation > 105 dB

Diagnosis > 85 dB

Electrode offset potential ±300mV

Leakage Current < 10 uA

Baseline Recovery < 3 S After Defi.

ECG Signal Range ±8 m V ( Vp-p )

Bandwidth Surgery 1 ~ 20 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

Calibration Signal 1 (mVp-p), Accuracy:5%

ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV

ARR Detecting Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,

PNP, PNC,Alarm Available,Review Available

Respiration

Method Impedance between R-F(RA-LL)

Differential Input Impedance > 2.5 MΩ

Measuring Impedance Range 0.3~3 Ω

Base line Impedance Range 200~4 KΩ

Bandwidth 0.1~2.5Hz

Resp. Rate Measuring and Alarm Range

Adult 0 ~ 120 Brpm

Neo/Ped 0 ~ 150 Brpm

Resolution 1 Brpm

Accuracy ±2 Brpm

Apnea Alarm 10 ~ 40 S

MSL vet patient monitor for sale EKONMP06

NIBP

Method Oscillometric

Mode Manual, Auto, STAT

Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)

Measuring Period in STAT Mode 5 Min

Pulse Rate Range 40 ~ 240 bpm

Alarm Type SYS, DIA, MEAN

Measuring and alarm range Adult Mode

SYS 40 ~ 270 mmHg

DIA 10 ~ 215 mmHg

MEAN 20 ~ 235 mmHg

Pediatric Mode

SYS 40 ~ 200 mmHg

DIA 10 ~ 150 mmHg

MEAN 20 ~ 165 mmHg

Neonatal Mode

SYS 40 ~ 135 mmHg

DIA 10 ~ 100 mmHg

MEAN 20 ~ 110 mmHg

Resolution 1mmHg

Accuracy Maximum Mean error ±5mmHg

Maximum Standard deviation ±8mmHg

Overpressure Protection Adult Mode 297±3 mmHg

Pediatric Mode 240±3 mmHg

Neonatal Mode 147±3 mmHg

SpO2

Measuring Range 0 ~ 100 %

Alarm Range 0 ~ 100 %

Resolution 1 %

Accuracy 70% ~ 100% 2 %

0% ~ 69% unspecified

Actualization intervalabout 1 Sec.

Alarm Delay 10 Sec.

Pulse Rate Measuring and Alarm Range 20~300 bpm

Resolution 1bpm

Accuracy ±3bpm

TEMP

Channel 2

Measuring and Alarm Range 0 ~ 50 ℃

Resolution 0.1 ℃

Accuracy ±0.2 ℃

Actualization intervalabout 1 Sec.

Average Time Constant < 10 Sec.

IBP

Label ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and alarm range ART 0 ~ 300 mmHg

PA -6 ~ 120 mmHg

CVP/RAP/LAP/ICP-10 ~ 40 mmHg

P1 / P2-10 ~ 300 mmHg

Press Sensor Sensitivity 5 uV/V/mmHg

Impedance 300-3000Ω

Resolution 1 mmHg

Accuracy ±2% or 1mmHg which is greater

Actualization interval about 1 secretary

Best vet patient monitor for sale EKONMP06

Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature

Optional: Nellcor SpO2, EtCO 2, IBP, Thermal Recorder

Cheap vet patient monitor for sale EKONMP06

MSL TEAM picture

Welcome to medicalequipment-msl.com,If you have any demand in Patient monitor. Please feel free to

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Lead Mode
5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)
HR and Alarm Range Adult 15 ~
300 bpm
Resolution
1 bpm
Leakage Current <
10 uA
Calibration Signal 1 (mVp-p), Accuracy
5%
ARR Detecting Type
ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS
Measuring Interval in AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)
Measuring Period in STAT Mode
5 Min
Pulse Rate Range 40 ~
240 bpm
Measuring Range 0 ~
100 %
Alarm Range 0 ~
100 %
Accuracy 70% ~
100% 2 %
Measuring and alarm range ART 0 ~
300 mmHg
Press Sensor Sensitivity
5 uV/V/mmHg
Accuracy ±2% or
1mmHg which is greater
Standard
ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional
Nellcor SpO2, EtCO 2, IBP, Thermal Recorder

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.

NEED A CONFIGURATION?

Discuss this product with our team

Share your application, destination, and required configuration for a project-specific response.

Contact Us