Lab Equipment

laboratory equipment ESR Analyzer esr blood test EKONPUC-A

MODEL / EKONPUC-A

The equipment uses the infrared obstacle detection method, scan the tested blood regularly, and automatically detect the sub-interface of red cells and plasma from time to time.

  • blood regularly, and automatically detect the sub-interface of red cells and
  • The instrument has highly automatic functions which can work at any time and continuously detect.
  • temperature; when the instrument detection time is 60 minutes, it can automatically calculate
  • c)Instrument can choose the ambient temperature compensation. When selecting the function
PRODUCT OVERVIEW

Product Details

Quick Details

The equipment uses the infrared obstacle detection method, scan the tested blood regularly, and automatically detect the sub-interface of red cells and plasma from time to time.

esr test

esr blood test

esr in blood

Specifications

laboratory equipment ESR Analyzer EKONPUC-A

esr test

esr blood test

esr in blood

ESR Analyzer EKONPUC-A

ESR Analyzer EKONPUC-A Specifications:

Number of test samples: 10 items at the same time

Max test speed: 20 samples/hour

Suited ESR tube: 1.6ml ESR tube(vacuum or normal)

Samples volume: 1.6 ml anticoagulant blood(blood 1.28ml+anticoagulant0.32ml),

the height of the inputting tube 60 mm

Accuracy range of sample: 50mm~64mm

Read resolution: ±0.2mm(westergren method result±1mm)

Interval of reading:1 minute

Test time:The test time is 30 minutes, equivalents to 1 hour

westergren method: The test time is 60 minutes, equivalents to 2 hour westergren method

Display:128×240 LCD screen

Power: 12V(SELV) by external power supply

Working principle and structure characteristic

The equipment uses the infrared obstacle detection method, scan the tested

blood regularly, and automatically detect the sub-interface of red cells and

plasma from time to time. The tested data will be converted into results of

Wechsler by computer processing. This can be displayed on the LCD screen

or printer and can also be transmitted through the serial interface to a computer.

The instrument makes up detection of the host (include infrared detection devices,

computer control systems, thermal printer) and the adaptive power supply (SELV).

The instrument has the large display apparatus; the interface is simple appearance

and has the film keyboard operation. The operation is simple and convenient.

The instrument has highly automatic functions which can work at any time and continuously detect.

The main performance indicators and the work of environmental requirements

1 Normal working conditions

a)Relative humidity:≤70%;

b)Away from strong sources of electromagnetic interference ;

c)Prevent dust and vibration, air mustn’t contain corrosive gases, pollution level 2 ;

d)Power:220×(1±10 %)V,50×(1 ±2 %)Hz;

e)Input power:40×(1+10%)VA;

f)Environment with good grounding, transient over-voltage category of Class Ⅱ ;

g)5-minute warm-up time boot.

2 Environmental conditions of storage and transportation

(a)Ambient temperature range : -20℃~+55℃;

(1) Relative humidity range: ≤93%;

(2) Atmospheric pressure range : 500hpa~1060hpa;

(3) No corrosive gas, a well-ventilated.

3 Basic parameters

a) Test a sample bit: Instrument detection of up to 10 samples at the same time;

b) The maximum detection rate: 20 samples / h;

c) Blood sedimentation tube: After the State Drug Administration of the registered

register vacuum,blood sedimentation tube φ8.3 * 120 vacuum tubes ( ordinary blood sedimentation tube)

d) Sample size : blood drawing 1.60ml+ anticoagulant agent 0.4ml;

e)The permissible error of adding the samples range:50mm ~ 64mm;

f)Instrument readings resolution : ±0.1mm (results of Wechsler±1mm);

g)Reading time interval :1 minute;

h) Testing Time: 30 minutes, which is equivalent to 1 hour Wechsler; the testing

time of equipment is 60 minutes, which is equivalent to two hours Wechsler;

i) Display: 128x240 dot matrix LCD screen

j) Power Supply: Instrument from an external 12V DC adapter power supply voltage (SELV);

a) Outside measurement :310 x 190 x 180 (mm);

l) Quality:4.8kg;

4 Technical requirements

a)Testing range:0—140 mm/h(results of Wechsler);

b)Testing relevance:Wechsler results of the linear correlation coefficient r≥0.98;

c)Testing repetitive: CV < 3% ;

5 Main function

a)Sample Adding Method : Insert the sample any time; detect the sample any time;

b)The output of the results:Display and print samples’ No. (serial number), Wechsler

results and the whole process of erythrocyte sedimentation rate curve, detection time and

temperature; when the instrument detection time is 60 minutes, it can automatically calculate,

display and print Katz index;

c)Instrument can choose the ambient temperature compensation. When selecting the function,

when the room temperature is 18 ℃ -30 ℃, it amends to 18 ℃ according to the blood sedimentation

calibration test results. (It can be selected in accordance with laboratory, the default of

equipment is no compensated);

d)Equipment can be stored not less than 150 samples of the test results, including the

sample number (serial number), erythrocyte sedimentation rate values and the dynamic curve;

e)There are power-off protection function, when a sudden power failure, automatic

detection has been completed the tested results.

f)Built-in thermal printer preparation equipment, it has automatic, manual printing options.

g)Operation interface: Film keyboard operation;

h)Communication Interface: RS-232 serial interface;

what will you get?

what will you get
TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Number of test samples
10 items at the same time
Max test speed
20 samples/hour
Suited ESR tube
1.6ml ESR tube(vacuum or normal)
Samples volume
1.6 ml anticoagulant blood(blood 1.28ml+anticoagulant0.32ml)
the height of the inputting tube
60 mm
Accuracy range of sample
50mm~64mm
Read resolution
±0.2mm(westergren method result±1mm)
Interval of reading
1 minute
Test time
The test time is 30 minutes, equivalents to 1 hour
westergren method
The test time is 60 minutes, equivalents to 2 hour westergren method
Display
128×240 LCD screen
Power
12V(SELV) by external power supply
a)Relative humidity
≤70%;
d)Power
220×(1±10 %)V,50×(1 ±2 %)Hz;
e)Input power
40×(1+10%)VA;
(a)Ambient temperature range
-20℃~+55℃;
(1) Relative humidity range
≤93%;
(2) Atmospheric pressure range
500hpa~1060hpa;

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.

NEED A CONFIGURATION?

Discuss this product with our team

Share your application, destination, and required configuration for a project-specific response.

Contact Us