Lab Equipment > Mindray analyzer

Mindray bc 30s Auto Hematology Analyzer

MODEL / Mindray-bc-30s

At Mindray we pride ourselves in our dedication and experience in developing better solutions for small labs. Our new line of 3-part hematology analyzers is the culmination of that efort. Compact yet powerful, full featured yet afordable, the BC-30s is what a 3-part analyzer should be.

  • 4 inch TFT Touch Screen
  • LAN Port supports HL7 protocol
  • Support bi-directional LIS
PRODUCT OVERVIEW

Product Details

Quick Details

mindray analyzer

BC-30S Hematology Analyzers

Specifications

What a 3-part should be

At Mindray we pride ourselves in our dedication and experience in developing better solutions for small labs. Our new line of 3-part hematology analyzers is the culmination of that efort. Compact yet powerful, full featured yet afordable, the BC-30s is what a 3-part analyzer should be.

mindray bc 30s

BC-30s

Auto Hematology Analyzer

Technical Specifications

Principles

Impedance method for WBC, RBC and PLT counting

Cyanide free reagent for hemoglobin test

Performance

Parameter Linearity Range Precision (CV %) Carryover

WBC(10 9 /L) 0-200 ≤3.5% (4.0-6.9) ≤0.5%

≤2.0% (7.0 -15.0)

RBC(10 12 /L) 0-8.00 ≤1.5% (3.5-6.5) ≤0.5%

HGB(g/L) 0-280 ≤1.5% (100 -180) ≤0.5%

MCV(fL) ≤1.0% (70 -110)

PLT(109/L) 0-4000 ≤5.0% (100 -149) ≤1.0%

≤4.0% (150 -500)

Parameters

21 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC,

HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT,

P-LCR,P-LCC

3 histograms for WBC, RBC and PLT

Reagent

M-30D DILUENT

M-30CFL LYSE

PROBE CLEANSER

Sample Volume

Prediluted mode 20μL

Whole blood mode 9 μL

Throughput

70 samples per hour

Display

10.4 inch TFT Touch Screen

mindray analyzer

Multi-language

Chinese, English, Spanish, Portuguese, Russian, French, Bahasa Indonesia

Data Storage Capacity

Up to 500,000 results including numeric and graphical information

Communication

LAN Port supports HL7 protocol

Support bi-directional LIS

Interface

4 USB port (for external printer, software upgrade, barcode reader,

WIFI adapter,keyboard and mouse), LAN port (1)

Printout

Thermal recorder, 50 mm wide paper, various printouts formats

External printer optional

Operating Environment

Temperature: 10°C~40°C

Humidity: 10%~90%

Air pressure: 70kPa~106kPa

Power Requirement

100V-240V

≤300VA

50Hz/60Hz

Dimension and Weight

Dimension: Depth(410 mm) x width(300 mm) x height(400 mm)

Weight: ≤20Kg

BC-30S Hematology Analyzers

Sample 1: BC-30s shows fags“Leucopenia”, ” Lymph decreased”,“Pancytopenia” which mean white blood cell decreased, the low number of Lymphocyte and decreased of leukocyte, erythrocyte and plate count. Meanwhile “R3” fag is also displayed. Two kinds of fag messages are both supported to ensure clinicians have better understanding of sample results.

Mindray BC-30s Auto Hematology Analyzer

Sample 2: Flag “Anemia” means that the sample has the possibility of anemia.

Sample 3: Flag “Throm- bopenia” indicator is shown together with Platelets low fag.

Hematology Analyzer BC-30S
TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

parameters
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC
Prediluted mode
20μL
Whole blood mode
9 μL
Temperature
10°C~40°C
Humidity
10%~90%
Air pressure
70kPa~106kPa
Dimension
Depth(410 mm) x width(300 mm) x height(400 mm)
Sample 1
BC-30s shows fags“Leucopenia” , ” Lymph decreased”,“Pancytopenia” which mean white blood cell decreased, the low number of Lymphocyte and decreased of leukocyte, erythrocyte and plate count. Meanwhile “R3” fag is also displayed. Two kinds of fag messages are both supported to ensure clinicians have better understanding of sample results.
Sample 2
Flag “Anemia” means that the sample has the possibility of anemia.

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.

NEED A CONFIGURATION?

Discuss this product with our team

Share your application, destination, and required configuration for a project-specific response.

Contact Us