Split beam UV-vis spectrophotometer EKONUV03
MODEL / EKONUV03Split beam UV-vis spectrophotometer EKONUV03. ?Quantitative analysis, automatically build standard curve.
- ?Quantitative analysis, automatically build standard curve.
- ?USB data interface with online dedicated PC workstation, with more powerful function and data processing capabilities.
- UV-vis spectrophotometer Function
- ?Standard automatic 4-position cell holder, optional seven-position or 16-position cell holder
Product Details
Quick Details
Function:
Photometric measurement, sample can be easily measured absorbance and transmittance at a specified wavelength
Quantitative analysis, automatically build standard curve.
Dynamic testing, enzyme kinetics reaction rate.
USB data interface with online dedicated PC workstation, with more powerful function and data processing capabilities.
Specifications
Split beam UV-vis spectrophotometer EKONUV03

Split beam UV-vis spectrophotometer EKONUV03
EKONUV03 split beam UV-Vis spectrophotometer can meet the different analytical testing requirements in organic chemistry, biochemistry, pharmaceutical analysis, food testing,medicine and health, environmental protection, life sciences etc fields of scientific research and production
UV-vis spectrophotometer Function:
Photometric measurement, sample can be easily measured absorbance and transmittance at a specified wavelength
Quantitative analysis, automatically build standard curve.
Dynamic testing, enzyme kinetics reaction rate.
USB data interface with online dedicated PC workstation, with more powerful function and data processing capabilities.
UV-vis spectrophotometer Performance feature
Standard automatic 4-position cell holder, optional seven-position or 16-position cell holder
Light source correction mechanism, the light source will be locating the best position when self-testing each time so that it can fix the offset focus position possibly caused by vibration during the transportation and the user don’t have to debugging light source after change it
With excellent structural design and precise split beam optical system, configurations of the host and software control
The key devices are made of imported components, which ensure the performance with high reliability
Open sample chamber design, a variety of optional accessories suitable for different applications, such as reflection sample holder, solid sample holder, water bath and auto sampler.
MSL TEAM picture
Welcome to medicalequipment-msl.com,If you have any demand in ultrasound machine.
Technical Specifications
Technical parameters from the currently available product information.
| Specifications | |
| (Models) | EKONUV03 |
| (Operation mode) | Host operating |
| Wavelength range | 190 - 1100nm |
| Spectral bandwidth | 1.8nm or 4nm optional |
| Optical system | Split beam, Czerny-Turner; Grating 1200line/mm |
| Wavelength Accuracy | ±0.5nm |
| Wavelength Repeatability | ±0.2nm |
| Wavelength Resolution | 0.1nm |
| Photometric Range | -0.3 to 3A; 0-200%T; -9999 to 9999C |
| Photometric Accuracy | ±0.002A(0~0.5A);±0.004A(0.5~1A);±0.3%T(0~100%T) |
| Photometric Repeatability | ±0.001A(0~0.5A);±0.002A(0.5~1A);±0.15%T(0~100%T) |
| Stray light | ≤ 0.05%T (220 nm,NaI;340 nm,NaNO2) |
| Baseline Flatness | ±0.002A(under PC on-line operating) |
| Stability | ±0.001A/hr @500nm,0A |
| Photometric noise | ±0.001A |
| Data output | 128×64 array graphic LCD |
| Power | 100W(kV:80V~250V) |
| Size | 470*400*230 |
| Weight | 16kg |
Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.
Discuss this product with our team
Share your application, destination, and required configuration for a project-specific response.